Updated:2026-05-21
Click Count:0 1. Preoperative preparation
Verify patient information, assess the condition, skin status, and contraindications, inform the patient of the purpose and procedure of the operation, and obtain informed consent.
Prepare consumables such as peripheral nerve stimulator, special stimulation needle, sterile dressing, iodophor, cotton swab, sterile gloves, etc.
Assist the patient to adopt a comfortable position, fully expose the operation site, relax the limbs, and fix the position to avoid shaking.
Turn off surrounding interference devices, debug the neurostimulator, and check the battery level, wiring, and integrity of the needle body.
II. Positioning and Disinfection
Based on anatomical landmarks and nerve course, accurately mark the puncture site of the target peripheral nerve.
Centered around the puncture point, apply iodophor disinfection to the skin in a circular pattern from the inside out, ensuring that the disinfection covers a diameter of ≥8cm. Allow the skin to dry naturally.
Operators wear sterile gloves and adhere to the principles of aseptic technique throughout the entire process.
III. Instrument parameter presetting
Turn on the nerve stimulator, with initial settings of current ranging from 0.5 to 1.0 mA, pulse frequency of 1 to 2 Hz, and pulse width of 0.1 to 0.3 ms.
Confirm that the positive and negative electrodes of the instrument are connected correctly, and connect the stimulation needle lead.
IV. Puncture and needle insertion
Hold the stimulating needle steadily and puncture the marked point, slowly advancing the needle layer by layer while avoiding blood vessels, tendons, and organs.
Observe the patient's sensation while inserting the needle, ensuring a steady insertion and avoiding sharp or deep punctures.
V. Precise neural localization
Slowly decrease the current of the stimulator and observe the contraction response of the corresponding muscle group in the limb.
When precise muscle twitching occurs and the patient experiences numbness, soreness, and radiating sensations, adjust the angle and depth of the needle slightly.
When the current drops to 0.2~0.4mA, there is still significant muscle contraction under neural innervation, indicating the presence of the optimal neural target site.
VI. Standardize stimulation therapy
Fix the needle position, lock the instrument parameters, and set the stimulation duration according to the diagnosis and treatment plan.
During treatment, closely observe the patient's facial expression, physical sensation, and limb reactions, and inquire about any discomfort at any time.
Adjust the current intensity slightly based on tolerance, and strictly avoid ultra-high current stimulation.
VII. End of treatment
First, set the stimulator current to zero, turn off the device power, and disconnect the connecting wires.
Press the puncture site aseptically and slowly withdraw the stimulation needle at a steady speed.
Disinfect the puncture site again, apply pressure to stop bleeding, and apply sterile dressing for protection.
VIII. Postoperative organization and education
Check and organize medical equipment and consumables, and dispose of disposable needles in accordance with medical waste disposal regulations.
Inform the patient of postoperative precautions: avoid contact with water for a short period of time, avoid strenuous activities, and pay attention to any abnormalities such as swelling, pain, or numbness in the local area.
Record the operation site, parameters, duration, and patient response, and improve the diagnosis and treatment ledger.
IX. Core operational taboos
It is strictly prohibited to perform puncture on areas with skin damage, infection, hematoma, or tumor
It is contraindicated for pregnant women, patients with severe arrhythmia, those wearing cardiac pacemakers, and individuals with coagulation dysfunction
Blind deep puncture in the spinal canal and near major blood vessels is strictly prohibited